Application:- Analysis technology- Safety circuits in vacuum systems- Control of ionization vacuum meters- General pressure measurement and control in systems in the fine and coarse vacuum range

Measurement principle: Heat conduction according to Pirani
Measurement range (Air, O2, CO, N2): 5×10^-4 … 1000 mbar
Accuracy (N2): 1×10^-3 … 100 mbar ±15% of the measured value
5×10^-4 … 1×10^-3 mbar ±50% of the measured value
100 … 1000 mbar ±50% of the measured value
Resolution: 1% of the measured value
Repeatability: 1×10^-3 … 100 mbar 2% of the measured value
Output signal (Measurement signal)
Voltage range: 0 … +10.3 VDC
Measurement range: +1.9 … +10.0 VDC
Voltage-pressure relationship: Logarithmic 1.286 V/decade
Error signal: 0 … +0.5 V (filament break)
Output impedance: 2×4.7
Minimum load impedance: 10 k, short-circuit proof
Response time: 80 ms
Transmitter identification: 27.0 k, against supply ground
Adjustment: One button for ATM and HV adjustment
Switching function: SP1, SP2
Threshold display and setting: One button for threshold output at the measurement signal output. Setting occurs with prolonged or repeated pressing
Setting range: 2×10^-3 … 500 mbar
Hysteresis: 10% above the lower threshold
Relay contact: 30 V, 0.5 ADC, potential-free
Closed: at low pressure (light on)
Open: at high pressure, error, missing supply

Condition Used
Age rating No age restriction
Model TTR 91
Manufacturer Leybold
Content 1 piece
Weight 250 g
Item ID 1234577188
Item ID 1234577188

Any further Questions? Just fill the form below and let us know how we can help you.

Leybold

Leybold Thermovac TTR 91 Transmitter without switching point DN 16 KF


Item number 230035
EAN
Condition Used
Available quantity
1
EUR 133.78 *
Content 1 piece
Unit price EUR 133.78 / piece
immediately available
* Excl. VAT excl. Shipping